o X-ray system must support both wired and 5Ghz wireless operation. o System must include high-definition Non-Glass DR panel: Approximate size 10” x 12” and 14” x 17” with 4mm or less margin. o System Software must be able to select and remotely control all Golden Engineering Pulse generators and Teledyne Constant Potential generators including the new LiteX. o System must have system software-controlled generator firing and sequencing programmed pulse, volley and delay interval capability. o DR Imager must have an operational kV Range of 40-400kV o DR Imager must have an operational high Temp. range of -21 - 150 degrees F sustained. o Must have wired and wireless controlled “Indefinite Timeframe Exposure Window”. o System must include seamless auto-mosaic “Auto-Stitch” image stitching capability. o Must have software-controlled radio frequency capability and easily integrate with robotic platforms. o All components must be easily user interchangeable with other systems in the field to include Calibration files. o System must meet the Ingress Protection rating of IP65 (Panel at IP67) for dust, dirt, splash; and a 4-foot drop shock tolerance. *(with no internal shock sensors) * o System must have Short-Range Tactical and Long-Range modes of wireless operation and have real-time Wi-Fi visual signal strength monitoring and selfhealing Wi-Fi connectivity. o DR Imager Panel must have Tungsten layer for Circuitry and Backscatter Protection. o System must support date, time and pulse count stamps on each image shot and Photo collection capability.