SPECIFICATIONS: 1. Must be able to simultaneously calibrate in one dispense up to 12 channels of a multi-channel pipette with a minimum volume of 0.5 ul and maximum volume of 1000ul. 2. Must include an ISO 8655 compliant multichannel balance with a readability of 0.1 mg and .01 mg. The multichannel balance must have an overload protection feature that warns the user when the weighed sample exceeds the capacity of the balance. 3. The system must include a microanalytical weighing table that ensures vibration-free operation and fast stabilization time 4. The microanalytical weighing table must be ergonomically built to allow height adjustments to prevent strain on user’s neck and back and support for the elbows during repetitive motions. 5. Software for managing pipette calibrations must be compatible with the Calibry system or equivalent. Must be able to connect to 5 systems. 6. Software must include user management, report release, change history, and audit trails. 7. Must include validation of the software to ensure proper operation. 8. Must include Installation Qualification (IQ)/Operational Qualifications (OQ) testing. 9. Must include at least one-year maintenance/warranty. 10. Total price must include multi-channel calibration system, software, ergonomic microanalytical weighing table, accessories, shipping, installation, training, warranty, two (2) years of extended care and taxes. The extended care shall include full preventive maintenance on-site, full repair coverage, and calibration