FTK 9519-4200 QuantiFit2 Respirator Fit Testing System with Bluetooth Specifications 23.2. Calibration: NIST Traceable Yes ___ No ___ 23.3. Input Voltage: 12VDC, Center Positive, 3A Max Yes ___ No ___ 23.4. Power Supply: Universal, 100-240VAC, 50/60Hz 1A Max Yes ___ No ___ 23.5. Operating Temperature: 5 to 45C Yes ___ No ___ 23.6. Humidity Range: 20 to 80%, Non-Condensing Yes ___ No ___ 23.7. Setup Memory: 20GB Yes ___ No ___ 23.8. Size: (H) 11 x (W) 4.4 x (D) 15 inches Yes ___ No ___ 23.9. Weight: 8.5 lbs Yes ___ No ___ 23.10. Touchscreen: Capacitive Touch (800 x 480 pixel display) Yes ___ No ___ 23.11. Dynamic Range 23.12. Leak Test Measurement: 2-5000 cc/min Yes ___ No ___ 23.13. Resolution: 0.1 cc/min Yes ___ No ___ 23.14. Fit Factor Computation: 6-53,000 Yes ___ No ___ 23.15. Resolution: 1 Yes ___ No ___ 23.16. Pressure Sensor-Fit Testing 23.17. Pressure Range: 0-20 in H2O Yes ___ No ___ 23.18. Resolution: 0.01 in H2O Yes ___ No ___ 23.19. Accuracy: ± 0.25% FS Yes ___ No ___ 23.20. Over Pressure Limit: 60 in H2O Yes ___ No ___ 23.21. Temperature Compensation: 15-30C Yes ___ No ___ 23.22. Pressure Sensor - Pump 23.23. Pressure Range: 0-20 PSI Yes ___ No ___ 23.24. Resolution: 0.1 PSI Yes ___ No ___ 23.25. Accuracy: ± 2.5% FS Yes ___ No ___ 23.26. Over Pressure Limit: 60 PSI Yes ___ No ___ 23.27. Temperature Compensation: 15-30C Yes ___ No ___ 23.28. Instrument Accuracy 23.29. Challenge Pressure: ± 5% Yes ___ No ___ 23.30. Leak Rate Measurement: ± 3% or ± 3 cc/min (whichever is greater) Yes ___ No ___ 23.31. USB Interface 23.32. 4Type A Ports (Keyboard, USB Memory, Printer) Yes ___ No ___ 23.33. 1 Type B Port (PC Communication)