1. VHA | EKHCS | SPS DEPARTMENT | WATER PURIFICATION TESTING AND PM&R
SERVICES
2. Contractor will provide, install, monitor, maintain and repair/replace, if necessary, a
critical water system capable of producing water that meets all AAMI ST108 standards
for critical and utility water utilized in SPS, following all applicable tables in ST108
(2,4,5,6)
3. Incoming water to both sites (Topeka and Leavenworth) are extremely different and will
require differing amounts of regular filter changes. It may be advisable to install a sand
type filter in the system at the Topeka site.
4. Contractor will provide a monthly written report showing continued compliance with
AAMI ST108 standards to the COR at each campus, Topeka, and Leavenworth.