This is a combined synopsis/solicitation for commercial products or commercial services prepared in accordance with the format in subpart 12.6, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; proposals are being requested and a written solicitation will not be issued.
DESCRIPTION:
Inspect and repair services for 7890B Gas Chromatograph (GC) with Flame Ionization Detector (FID) to bring equipment to a functional state.
Functional state is defined by the following:
• FID flame is able to light.
• FID signal reads 0 Hz when the detector’s electrometer is turned off.
• FID signal increases when the detector’s electrometer is turned on.
• FID signal changes in response to changes in the flow of gases applied to the FID (real-time responsiveness to applied stimuli)
The below parts will be installed into the GC/FID if the currently installed part is determined to be defective.
7890B GC/FID parts:
- FID interface board
- FID electrometer PCA
- FID interconnect assembly
- FID collector assembly
The Vendor is to provide a quote (Including parts and labor) for any additional recommended repairs.