Please see attached RFQ and Specification documents for details on this opportunity.
March 31, 2025, Amendment 0001
To answer received question from Industry:
Q: Could you provide the type of calibration needed for the Delta DAQ F/T Transducer with IP68 protection sensor?
A: The calibration will need to define the measurement range as defined by ATI US-150-600, and a calibration matrix must be provided to convert voltages to loads. The required sensing ranges are Fx,Fy=150lbf, Fz=450lbf, Tx,Ty,Tz=600in-lbf. All 6 axes must be defined and calibrated to accurately achieve the requested measurement range.
April 14, 2025- AMENDMENT 0002
The purpose of this amendment is to add the calibration required to the minimum specifications document. As a result of this, the response date has been extended to Wednesday, April 16, 2025.