Specifications include, but are not limited to: A high vector magnetic field, low vibrating closed cycle measurement system optimized for confocal microscope measurements 1. Closed cycle liquid-helium free cryostat to guarantee operation independent from liquid helium logistics. 2. The base temperature of the cryostat is less than 1.7 Kelvin. 3. The minimum Z-direction magnetic field is 9 T. 4. The minimum X-direction and X-direction magnetic field is 1 T and 1 T. 5. Top or bottom loading module to allow fast sample exchange while maintaining the superconducting magnet cold and operational, and the maximum sample exchange time is six hours. 6. Low-vibration cryostat suitable for scanning probe microscope measurements. 7. Low drift cryostat under a high magnetic field. The maximum drift is less than 1 um under a 9 T magnetic field. 8. The objective for confocal microscope measurements is at the same temperature as the cryostat to ensure the sample temperature is unaffected by thermal radiation. The sample temperature should be less than 1.7 Kelvin. 9. XYZ-direction nano positioner, non-magnetic, temperature range (1K-400K), minimum travel range is 5 mm. 10. XY-direction scanner for high precision scanning, non-magnetic, temperature range (1K-400K), minimum scan range is 125 um at base temperature.