Specifications include, but are not limited to: MANUFACTURERS A. Basis of Design: Steelco Model _VS263639/1 LE WE ECO CD Single door, Cabinet style autoclave. B. Acceptable manufacturers a. Steelco b. Steris c. Getinge PART 2-Dimensions Interior Chamber Capacity Overall Dimension 26”W x 36”H x 39”D 21.5 ft3 53.5”W x 73.5”H” x 53”D” PART 3-Sterilizer Construction Sterilizer and all parts to be new, remanufactured units not acceptable The vessel shall be manufactured in the ASME code factory of the named sterilizer manufacturer. Use of Non-proprietary parts is required Vessel shall be rectilinear in shape and include a fully jacketed chamber. Chamber jacket and door materials shall be non-laminated materials, solid stainless-steel plate or sheet. Chamber and doors must be type 316L stainless steel. Nickel clad not acceptable Chamber structural channel supports shall extend beyond the chamber radiused corners, shall be mitered and seam welded at their intersections. Chamber, back head, headring and door shall be solid 316L stainless steel Jacket shall be type 304 stainless steel. Frame and vessel support structure shall be welded steel frame, painted. All internal vessel structural support including doors, shall be type 316L stainless steel. Longitudinal welds and adjacent surfaces shall be finished to a porosity to <10” Ra. Insulation shall be min. 1” inch thick rigid fiberglass, protected by an .063” aluminum sheet covering. All pressure containment areas shall be designed for 45 PSIG at full vacuum. Steam baffles shall be positioned in the chamber to minimize wetting by condensate and insure proper steam distribution within the chamber. Stainless Steel boiler with automatic boiler blowdown Heating Element Service on Right Side