Specifications include, but are not limited to: 2.2.1 The system must allow for walk-away operation with minimal operator monitoring or intervention. 2.2.2 Process samples at a rate of up to 210 samples per hour. 2.2.3 XYZ autosampler with up to 240 positions 2.2.4 The System must be able to support additional modules for different chemistries; each with its own power supply, light source, and detector. Each module must be a standalone unit without any additional cost. 2.2.5 System must allow full view of fluid movement in mixing manifold. 2.2.6 Display the full absorbance spectrum in real time. 2.2.7 All methods must include manufacturer lab tested data collected using that method and specified instrumentation. Data to include linearity, accuracy, precision, method detection and reporting limit across multiple days. 2.2.8 All chemistries must use US EPA or ISO accepted methodologies. 2.2.9 Must include a heating system to maintain it at a constant temperature. 2.2.10 Small analyzer footprint. 2.2.11 Flow cell with up to 100 mm path length. 2.2.12 System must use an array spectrometer detector, allowing utilization of alternative wavelengths to extend the linear range. 2.2.13 Computer with a monitor to run instrument software. 2.2.14 Start up consumables/maintenance kit including spare tubing, tools etc.