Specifications include, but are not limited to: o Line Item #1 Fall: 9-1-24 to 12-31-24 – Counties - Honolulu, Maui, Kauai and Hawaii o Line Item #2: Spring 3-1-25 to 5-1-25 - County - Honolulu Calibrating technician shall follow manufacturers’ instructions and be familiar with the equipment and controls. Equipment shall be verified according to the latest updates to AASHTO R18 & R28 and calibrated as needed. Quotation is for the verification and calibration of all equipment at the lab and field office locations listed on the attachments. Calibration certification is for a period of 1 year unless noted otherwise. Documentation shall include calibration reports and traceability of standards used. Deliver Report/Certificate of Calibration/Verification including: a. Name of technician performing calibration/verification. b. Date calibration/verification was performed. c. Equipment and/or standards used to perform calibration/verification. d. Procedures used in calibration/verification. e. Test data from calibration/verification. f. Description of any corrective measures performed. Special Requirements for Satec test machine as listed on the attachment Oahu Honolulu Central Lab: Experience in using Satec/Instron Nu Vision Partner Software test procedure or a Satec/Instron approved program is required to operate the machine's closed loop controls during the verification. During the verification procedure, the spherical head shall be protected with a full plate to prevent any damage from the load-verifying device. Load verifications shall be performed throughout the digital output range. Provide document demonstrating calibration standard traceable to NIST.