• The system should have compact bench-top design occupying a footprint of < 0.5 m2 to fit the space constraints of the laboratory and be of open architecture sample introduction area. • The mainframe must not weigh more than 100 kg. • The system must be equipped with a collision reaction cell. • The instrument should be delivered with all necessary supplies and accessories required for the installation and start-up. The installation and training must be included. Sample Introduction • 3-channels computer controlled peristaltic pump for pumping of sample, internal standard and spray chamber drain. • Thermoelectrically controlled quartz Scott type spray chamber with concentric nebulizer. • The proposal must include the capability of analyzing, with its standard sample introduction system, high percent levels of dissolved solids (up to ~25% TDS) automatically without any manual, automated or on-line dilution of the sample with varying pump tubing sizes...