2.1.Ability to perform an automated dual solid phase extraction and pressurized fluid extraction of any solid or semi-solid sample type and include a pre-programmed library of methods. 2.2.Providing a filtered, room temperature extract in as little as 15 minutes. The extract must be dispensed at room temperature to allow for immediately processing. 2.3.Ability to process a minimum of 12 samples in series in as little as 3 hours with the capability to program up to 24 samples at a time. 2.4.One extraction vessel capability of running both smaller and larger sample sizes. Eliminating the need to stock multiple vessel types. 2.5.The ability to add sorbents and cleanup components such as salt, C18, carbon black, sand and diatomaceous earth for in situ sample extraction and cleanup. 2.6.The ability to run extractions in a pressurized environment such that the solvent doesn’t boil. 2.7. Can run up to 12 cycles on a single sample. 2.8.A solvent array capable of up to 6 solvents being loaded on the system at a time. Capatible with a wide range of organic and aqueous solvents and solvent mixtures. 2.9. Control by an integrated touch screen controller, which includes a built-in video library including training videos. It must not require the use of any external controllers such as terminals, tablet or PC device. 2.10. Extractions whould be cooled during the extraction process in a timely manner using an integrated cooling coil and must not require additional apparatus wuch as water baths, chillers or waiting period.