The successful proposer shall provide a service that: 1. Provides an FDA approved Cardiac Monitor in the Pre-Hospital setting with a minimum of the following requirements. A. Electrocardiogram Monitoring a. Capable of providing basic cardiac rhythm monitoring b. Capable of providing diagnostic quality 12-lead ECG c. Display screen that shows current Heart Rate, Blood Pressure, Respiratory Rate, SPO2, Pleth Waveform, ECG waveforms, ETCO2 waveform and numeric value d. Current time display e. User performed self-test to include energy delivery test f. Data transmission capability via internal or external Modem (minimum of WiFi capability) g. All cables and supplies: i. Limb lead cable ii. 12-lead cable iii. SPO2 Cable and Finger Probe iv. Carbon monoxide Cable and Finger Probe v. Electrical therapy cable B. Carrying case with applicable storage compartments, shoulder strap, and handle for each monitor C. External pacer a. Continuation of cardiac pacing if limb lead is disconnected b. Provide pacing with user selected rate and mA c. Pace via handsfree pads d. Allow for underlying rhythm interpretation...