Desired Autoclave Sterilizer Specifications: Encore 20, Inside chamber dimensions: 20”W x 20”H x 38”L, Volume: 8.8 cu ft. / 249 liters, Single door vessel with vertical slide door, Certified used pressure valve, Vessel is hydrostatically and ultrasound tested, Patented Water-Mizer water conservation package included, Recessed or Cabinet mounted, Steam source required, Encore 2G Control System, Mitsubishi programmable logic control (PLC), 8.4” high resolution LCD color touch screen display, Twenty (20) user configurable cycles (115-135C), Two (2) user configurable low temp cycles (100-114C), Factory configured leak test and Bow Dick (air removal test), Controlled slow exhaust for liquid cycles, Panel mount impact printer for cycle documentation, Data logging to removable USB drive for cycle documentation, Graph/Data display of current cycles, Historical data display of past cycles, Alarm history display, Multi-layered configurable password security